Strany 1 - Operation Manual

Operation Manualfor XP2i Digital Test Gauge (Standard and Dual-Display(DD))

Strany 2 - Contents

Functions  8XP2i Operation ManualResetIf for some reason the XP2i needs to be reset, remove any battery for at least one minute, then reinstall the

Strany 3 - Overview

Functions  9XP2i Operation ManualInternal Data Logging (-DL DataLoggerXP Option)With the DataLoggerXP firmware update option you can record up to six

Strany 4 - OPERATING INSTRUCTIONS

Enclosure  10XP2i Operation ManualEnclosureAluminum alloy, designed to meet IP67 ratings.Weight: 562 g (19.8 oz), including batteries. ALL DIMENSIONS

Strany 5 - Functions

Functions  11XP2i Operation ManualXP2I SERIAL NUMBERSSerial Number LocationThe serial number of your XP2i is located on the stem above the pressur

Strany 6 - Functions  4

Specifications  12XP2i Operation ManualSpecificationsAccuracyIncludes all effects of linearity, hysteresis, repeatability, temperature, and stability fo

Strany 7 - (peak) button

Specifications  13XP2i Operation ManualMedia CompatibilityLiquids and gases compatible with 316 Stainless Steel.Pressure Conversions1 PSI = 27.6806 i

Strany 8 - Measuring Vacuum

Specifications  14XP2i Operation ManualPOWERBatteries . . . . . . . . . . . . . . . . . . . . . . . . Three size AA (LR6) batteries. !WARNING: Do no

Strany 9 - (on/o) b

Specifications  15XP2i Operation ManualPRESSURE RANGES, DISPLAY SCALES, & RESOLUTIONPSI bar kPa Overpressure PSIinH20 inHg mmHg mmH20 kg/cm2bar

Strany 10 - PSVtest Mode

Safety & Certifications  16XP2i Operation ManualSafety & CertificationsHAZARDOUS LOCATIONSEvery XP2i pressure gauge includes the following In

Strany 11 - Communications/Programming

Safety & Certifications  17XP2i Operation ManualEMC EC Declarations of Conformity ATEX EC Declarations of ConformityEC Declaration of Conformity I

Strany 12 - Enclosure

ContentsOverview ... 1Introduction ...

Strany 13 - The serial number is located

Safety & Certifications  18XP2i Operation ManualPED EC Declarations of ConformityI/We of declare that EC Declaration of Conformity Crystal Enginee

Strany 14 - Specifications

Safety & Certifications  19XP2i Operation ManualATEX/IECEX SAFETY INSTRUCTIONSBezpečnostní Instrukce Pro Prostředí s Nebezpečím Výbuchu – ČESKY

Strany 15 - Connections

Safety & Certifications  20XP2i Operation ManualSafety Instructions for Hazardous Locations – ENGLISH (English)• Do not use the RS-232 connector

Strany 16 - XP2i Operation Manual

Safety & Certifications  21XP2i Operation ManualInstructions De Sécurité Pour Les Zones Dangereuses – FRANÇAIS (French)• Ne pas utilisez le conne

Strany 17 - MODEL NUMBERING SYSTEM

Safety & Certifications  22XP2i Operation ManualVeiligheidsinstructie Voor Gebruik In Een Explosie Gevaarlijke Omgeving – NEDERLANDS (Dutch)• Het

Strany 18 - Safety & Certifications

Safety & Certifications  23XP2i Operation ManualRäjähdysvaarallisten Tilojen Turvallisuusohjeita – SUOMEN KIELI (Finnish)• RS-232 väylää/liitintä

Strany 19 - EC Declaration of Conformity

Support  24XP2i Operation ManualSupportTROUBLESHOOTINGThe XP2i is a very high performance gauge. Due to the high resolution of this product, you m

Strany 20

Support  25XP2i Operation ManualCALIBRATIONIf adjustment is required, we recommend returning the unit to the factory. Factory service offers benefit

Strany 21 - DEUTSCH (German)

Support  26XP2i Operation ManualSOFTWARELabVIEW™ Drivers and Programming InstructionsControl and communicate with XP2is using National Instrument’

Strany 22 - ESPAÑOL (Spanish)

Support  27XP2i Operation ManualCONTACT USPhone . . . . . . . . . . . . . . . . . . . . . . . . . .(805) 595-5477Toll-Free . . . . . . . . . . .

Strany 23 - ITALIANO (Italian)

Overview  1XP2i Operation ManualOverviewINTRODUCTIONThank you for choosing the XP2i Digital Test Gauge from Crystal Engineering Corporation. Your

Strany 24 - POLSKI (Polish)

4661.C1© 2013 Crystal Engineering Corporation708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701

Strany 25 - SUOMEN KIELI (Finnish)

Overview  2XP2i Operation ManualOPERATING INSTRUCTIONSThe XP2i is shipped with batteries installed, so it’s ready to use. Press and hold the (on/o

Strany 26 - TROUBLESHOOTING

Functions  3XP2i Operation ManualFunctions Units ButtonPressing this button causes the XP2i to select the next available unit of pressure measureme

Strany 27 - Support  25

Functions  4XP2i Operation ManualTare (-DD, Dual-Line Display XP2i Only)Tare is a constant value subtracted from the true pressure. For instance, if

Strany 28 - ACCESSORIES

Functions  5XP2i Operation Manual Peak Detection and AverageOn the XP2i, pressing the Peak button causes the display to cycle through the following:

Strany 29 - WARRANTY

Functions  6XP2i Operation ManualAutomatic Shut-oThe XP2i has a shutoff timer and will turn off automatically after 20 minutes of non-operation. Pres

Strany 30

Functions  7XP2i Operation ManualWater Density (Inches of Water)The following applies only to models where inch of water is a selectable pressure un