TROUBLESHOOTING
Unstable Pressure
X
Problem: Pressure cannot be generated correctly, or set pressure does not remain stable.
X
Solutions:
Check that all adapters and pressure ttings have been tightened suciently to eliminate leaks.
Conrm test uid does not contain air (air in the uid can increase adiabatic eects). Replace test uid if necessary.
Replace hose (hose expansion can cause unstable pressure).
Pressure does not Increase
X
Problem: The system fails to indicate a pressure increase after considerable pumping action of the handle.
X
Solutions:
Close the pressure vent valve.
Remove the ll plug, and push the middle of the Check Valve In. If the vent valve cannot be reached because it’s not aligned with the ll plug,
loosen the reservoir lid and rotate until aligned.
Check that all adapters and pressure ttings have been tightened suciently.
Pump requires service (please contact your local sales oce for advice).

MAINTENANCE
Users should/must carry out the following cleaning procedures as and when required:
Exterior cleaning . . . . . . . . . . . . . . . . Clean using a soft, damp cloth.
Visual Inspection . . . . . . . . . . . . . . . . Make sure that the bottom of the reservoir and the uid does not contain contaminants or solids.