TROUBLESHOOTING
Unstable Pressure
X
Problem: Pressure cannot be generated correctly, or set pressure does not remain stable.
X
Solutions:
• Check that all adapters and pressure ttings have been tightened suciently to eliminate leaks.
• Conrm test uid does not contain air (air in the uid can increase adiabatic eects). Replace test uid if necessary.
• Replace hose (hose expansion can cause unstable pressure).
Pressure does not Increase
X
Problem: The system fails to indicate a pressure increase after considerable pumping action of the handle.
X
Solutions:
• Close the pressure vent valve.
• Remove the ll plug, and push the middle of the Check Valve In. If the vent valve cannot be reached because it’s not aligned with the ll plug,
loosen the reservoir lid and rotate until aligned.
• Check that all adapters and pressure ttings have been tightened suciently.
• Pump requires service (please contact your local sales oce for advice).
MAINTENANCE
Users should/must carry out the following cleaning procedures as and when required:
Exterior cleaning . . . . . . . . . . . . . . . . Clean using a soft, damp cloth.
Visual Inspection . . . . . . . . . . . . . . . . Make sure that the bottom of the reservoir and the uid does not contain contaminants or solids.